Applied sciences

Metrology and Measurement Systems

Content

Metrology and Measurement Systems | 2015 | vol. 22 | No 1

Download PDF Download RIS Download Bibtex

Abstract

An array consisting of four commercial gas sensors with target specifications for hydrocarbons, ammonia, alcohol, explosive gases has been constructed and tested. The sensors in the array operate in the dynamic mode upon the temperature modulation from 350°C to 500°C. Changes in the sensor operating temperature lead to distinct resistance responses affected by the gas type, its concentration and the humidity level. The measurements are performed upon various hydrogen (17-3000 ppm), methane (167-3000 ppm) and propane (167-3000 ppm) concentrations at relative humidity levels of 0-75%RH. The measured dynamic response signals are further processed with the Discrete Fourier Transform. Absolute values of the dc component and the first five harmonics of each sensor are analysed by a feed-forward back-propagation neural network. The ultimate aim of this research is to achieve a reliable hydrogen detection despite an interference of the humidity and residual gases.
Go to article

Authors and Affiliations

Patryk Gwiżdż
Andrzej Brudnik
Katarzyna Zakrzewska
Download PDF Download RIS Download Bibtex

Abstract

As it contains elements of complete digital impedance meter, the AD5933 integrated circuit is an interesting solution for impedance measurements. However, its use for measurements in a wide range of impedances and frequencies requires an additional digital and analogue circuitry. This paper presents the design and performance of a simple impedance meter based on the AD5933 IC. Apart from the AD5933 IC it consists of a clock generator with a programmable prescaler, a novel DC offset canceller for the excitation signal based on peak detectors and a current to voltage converter with switchable conversion ratios. The authors proposed a simple method for choosing the measurement frequency to minimalize errors resulting from the spectral leakage and distortion caused by a lack of an anti-aliasing filter in the DDS generator. Additionally, a novel method for the AD5933 IC calibration was proposed. It consists in a mathematical compensation of the systematic error occurring in the argument of the value returned from the AD5933 IC as a result. The performance of the whole system is demonstrated in an exemplary measurement.
Go to article

Authors and Affiliations

Konrad Chabowski
Tomasz Piasecki
Andrzej Dzierka
Karol Nitsch
Download PDF Download RIS Download Bibtex

Abstract

Ambient concentrations of CO, as well as NOx and O3, were measured as a part of the PARADE campaign conducted at the Taunus Observatory on the summit of the Kleiner Feldberg between the 8th of August and 9th of September 2011. These measurements were made in an effort to provide insight into the characteristics of the effects of both biogenic and anthropogenic emissions on atmospheric chemistry in the rural south-western German environment. The overall average CO concentration was found to be 100.3±18.1 ppbv (within the range of 71 to 180 ppbv), determined from 10-min averages during the summer season. The background CO concentration was estimated to be ~90 ppbv. CO and NOx showed bimodal diurnal variations with peaks in the late morning (10:00-12:00 UTC) and in the late afternoon (17:00-20:00 UTC). Strong correlations between CO and NOx indicated that vehicular emission was the major contributor to the notable CO plumes observed at the sampling site. Both local meteorology and backward trajectory analyses suggest that CO plumes were associated with anthropogenically polluted air masses transferred by an advection to the site from densely populated city sites. Furthermore, a good linear correlation of R2 = 0.54 between CO and O3 (∆O3/∆CO=0.560±0.016 ppbv/ppbv) was observed, in good agreement with previous observations
Go to article

Authors and Affiliations

Jingsong Li
Andreas Reiffs
Uwe Parchatka
Horst Fischer
Download PDF Download RIS Download Bibtex

Abstract

To reduce the influence of the static unbalance on an infrared missile guidance system, a new static unbalance measure system for the gimbals axes has been developed. Considering the coupling effects caused by a mass eccentricity, the static balance condition and measure sequence for each gimbal axis are obtained. A novel static unbalance test approach is proposed after analyzing the dynamic model of the measured gimbal axis. This approach is to drive the measured gimbal axis to do sinusoidal reciprocating motion in a small angle and collect its drive currents in real time. Then the static unbalance of the measured gimbal axis can be obtained by the current multi-cycle integration. Also a measuring system using the proposed approach has been developed. A balanced simulator is used to verify the proposed approach by the load and repeatability tests. The results show the proposed approach enhances the efficiency of the static unbalance measurement, and the developed measuring system is able to achieve a high precision with a greater stability.
Go to article

Authors and Affiliations

Hui Yang
Yan Zhao
Min Li
Falin Wu
Download PDF Download RIS Download Bibtex

Abstract

The paper presents a technique for measuring membrane displacements with one motionless camera. The method consists in measuring the distance to an object based on one image obtained from a motionless camera with a fixed-focus lens. The essence of the proposed measurement technique is to determine changes of the distance between a membrane and a video camera based on analysis of changes in the focus view of a marker placed on the membrane plane. It is proven that the used technique allows to monitor the frequency and amplitude of the membrane vibration. The tests were performed for the oscillation frequency in the range from 0.5 Hz to 6 Hz and deviations from the neutral position in the range of ±3 mm.
Go to article

Authors and Affiliations

Krzysztof Murawski
Download PDF Download RIS Download Bibtex

Abstract

This paper presents a new test method able to infer - in periods of less than 7 seconds - the refrigeration capacity of a compressor used in thermal machines, which represents a time reduction of approximately 99.95% related to the standardized traditional methods. The method was developed aiming at its application on compressor manufacture lines and on 100% of the units produced. Artificial neural networks (ANNs) were used to establish a model able to infer the refrigeration capacity based on the data collected directly on the production line. The proposed method does not make use of refrigeration systems and also does not require using the compressor oil.
Go to article

Authors and Affiliations

Rodrigo Coral
Carlos A. Flesch
Cesar A. Penz
Maikon R. Borges
Download PDF Download RIS Download Bibtex

Abstract

This paper deals with the amplitude estimation in the frequency domain of low-level sine waves, i.e. sine waves spanning a small number of quantization steps of an analog-to-digital converter. This is a quite common condition for high-speed low-resolution converters. A digitized sine wave is transformed into the frequency domain through the discrete Fourier transform. The error in the amplitude estimate is treated as a random variable since the offset and the phase of the sine wave are usually unknown. Therefore, the estimate is characterized by its standard deviation. The proposed model evaluates properly such a standard deviation by treating the quantization with a Fourier series approach. On the other hand, it is shown that the conventional noise model of quantization would lead to a large underestimation of the error standard deviation. The effects of measurement parameters, such as the number of samples and a kind of the time window, are also investigated. Finally, a threshold for the additive noise is provided as the boundary for validity of the two quantization models
Go to article

Authors and Affiliations

Diego Bellan
Download PDF Download RIS Download Bibtex

Abstract

The vapour pressure of most explosives is very low. Therefore, the explosive trace detection is very difficult. To overcome the problem, concentration units can be applied. At the Institute of Optoelectronics MUT, an explosive vapour concentration and decomposition unit to operate with an optoelectronic sensor of nitrogen dioxide has been developed. This unit provides an adsorption of explosive vapours from the analysed air and then their thermal decomposition. The thermal decomposition is mainly a chemical reaction, which consists in breaking up compounds into two or more simple compounds or elements. During the heating process most explosive particles, based on nitro aromatics and alkyl nitrate, release NO2 molecules and other products of pyrolysis. In this paper, the most common methods for the NO2 detection were presented. Also, an application of the concentration and decomposition unit in the NO2 optoelectronic sensor has been discussed.
Go to article

Authors and Affiliations

Beata Zakrzewska
Download PDF Download RIS Download Bibtex

Abstract

A sensing system utilizing a standard optical fiber as a distributed sensor for the detection and localization of mechanical vibrations is presented. Vibrations can be caused by various external factors, like moving people, cars, trains, and other objects producing mechanical vibrations that are sensed by a fiber. In our laboratory we have designed a sensing system based on the Φ-OTDR (phase sensitive Optical Time Domain Reflectometry) using an extremely narrow laser and EDFAs.
Go to article

Authors and Affiliations

Radim Sifta
Petr Munster
Petr Sysel
Tomas Horvath
Vit Novotny
Ondrej Krajsa
Miloslav Filka
Download PDF Download RIS Download Bibtex

Abstract

The paper presents a study of the performance of some selected UV detectors. Unlike many similar works, the obtained data refer to commercial photodiodes (not only to detector materials). The main task of the research was to determine the influence of the operating temperature and annealing on the detector spectral responsiveness. A comparison of the results obtained for the photodiodes made of GaN and SiC was also performed. Although both kinds of detectors can work at high temperatures for a long time, some modification of their properties was observed. However, for GaN and SiC photodiodes, this modification has a substantially different nature. It is very important for some applications, e.g. fire alarms and a military equipment.
Go to article

Authors and Affiliations

Joanna Ćwirko
Robert Ćwirko
Janusz Mikołajczyk
Download PDF Download RIS Download Bibtex

Abstract

The aim of this study was to assess the innovation risk for an additive manufacturing process. The analysis was based on the results of static tensile tests obtained for specimens made of photocured resin. The assessment involved analyzing the measurement uncertainty by applying the FMEA method. The structure of the causes and effects of the discrepancies was illustrated using the Ishikawa diagram. The risk priority numbers were calculated. The uncertainty of the tensile test measurement was determined for three printing orientations. The results suggest that the material used to fabricate the tensile specimens shows clear anisotropy of the properties in relation to the printing direction.
Go to article

Authors and Affiliations

Stanisław Adamczak
Jerzy Bochnia
Bożena Kaczmarska
Download PDF Download RIS Download Bibtex

Abstract

Photovoltaic (PV) power optimizers are introduced in PV systems to improve their energetic productivity in presence of mismatching phenomena and not uniform operating conditions. Commercially available converters are characterized by different DC-DC topologies. A promising one is the boost topology with its different versions. It is characterized by its circuital simplicity, few devices and high efficiency values - necessary features for a Distributed Maximum Power Point Tracking (DMPPT) converter. PV power optimizer designs represent a challenging task since they operate in continuously changing operating conditions which strongly influence electronic component properties and thus the performance of complete converters. An aspect to carefully analyze in such applications is the thermal factor. In this paper, a necessity to have a suitable temperature monitoring system to avoid dangerous conditions is underlined In addition, another important requirement for a PV power optimizer is its reliability, since it can suggest a useful information on its diagnostic aspects, maintenance and investments. In fact, a reliable device requires less maintenance services, also improving the economic aspect. The evaluation of the electronic system reliability can be carried out using different reliability prediction models. In this paper, reliability indices, such as the Mean Time Between Failure (MTBF) or the Failure Rate of a Diode Rectification (DR) boost, are calculated using the evaluation of the Military Handbook 217F and Siemens SN29500 prediction models. With the reliability prediction results it has been possible to identify the most critical components of a DMPPT converter and a measurement setup has been developed in order to monitor the component stress level on the temperature, power, voltage, current, and energy in the DMPPT design phase avoiding the occurrence of a failure that might decrease the service life of the equipment.
Go to article

Authors and Affiliations

Marcantonio Catelani
Lorenzo Ciani
Giorgio Graditi
Giovanna Adinolfi
Download PDF Download RIS Download Bibtex

Abstract

BC (Black Carbon), which can be found in the atmosphere, is characterized by a large value of the imaginary part of the complex refractive index and, therefore, might have an impact on the global warming effect. To study the interaction of BC with light often computer simulations are used. One of the methods, which are capable of performing light scattering simulations by any shape, is DDA (Discrete Dipole Approximation). In this work its accuracy was estimated in respect to BC structures using the latest stable version of the ADDA (vr. 1.2) algorithm. As the reference algorithm the GMM (Generalized Multiparticle Mie-Solution) code was used. The study shows that the number of volume elements (dipoles) is the main parameter that defines the quality of results. However, they can be improved by a proper polarizability expression. The most accurate, and least time consuming, simulations were observed for IGT_SO. When an aggregate consists of particles composed of ca. 750 volume elements (dipoles), the averaged relative extinction error should not exceed ca. 4.5%.
Go to article

Authors and Affiliations

Krzysztof Skorupski
Download PDF Download RIS Download Bibtex

Abstract

An active beam-pointing stabilization system has been developed for a high-power KrF laser system to eliminate the long-term drift of the directional change of the beam in order to have a stable focusing to a high intensity. The control of the beam direction was achieved by a motor-driven mirror activated by an electric signal obtained by monitoring the position of the focus of the output beam. Instead of large sized UV-sensitive position sensitive detectors a simple arrangement with scatter plates and photodiodes are used to measure the directionality of the beam. After the beam stabilization the long-term residual deviation of the laser shots is ~14 μrad, which is comparable to the shot-to-shot variation of the beam (~12 μrad). This deviation is small enough to keep the focal spot size in a micrometer range when tightly focusing the beam using off-axis parabolic mirrors.
Go to article

Authors and Affiliations

A. Barna
I. B. Földes
J. Bohus
S. Szatmári

Instructions for authors



Sample article with Author guidelines



Author guidelines



Types of contributions

Metrology and Measurement Systems welcomes submissions of the following article types:

• invited special issue or review papers presenting the current stage of the knowledge within scope of the journal (about 20 edited pages, approximately 3000 characters each),
• research papers reporting high-quality original scientific or technological advancements (max. 12 pages),
• papers based on extended and updated contributions presented at scientific conferences (max. 12 pages),
• short notes, i.e. book reviews, conference reports, short news (max. 2 pages).


Manuscript preparation

General The text of a manuscript should be written in clear and concise English. The camera-ready format – with attached separate files containing illustrations, tables and photographs – is required. A cover letter with clear explanation of scientific novelty of the paper is strongly recommended. Papers based on extended and updated contributions presented at scientific conferences, or strongly related to previous authors’ works, must be accompanied with a cover letter file, which should explain in details changes made in the manuscript in comparison with the original conference paper and highlight the novelty in reference to other authors’ works.
The main text of a manuscript should be printed on an A4 page (with margins of 2.5 cm) using Times New Roman style with a font size of 12 pt; the paragraphs should start with the indentation of 5 mm, and titles should be written in bold. That text can be divided into sections (numbered 1, 2, …), first-order subsections (numbered 1.1., 1.2., …, written in italics), and – if needed – second-order subsections (numbered 1.1.1., 1.1.2., …, written same as first-order subsections). The only acceptable manuscript formats are in Microsoft Word (.doc, .docx).

The Editor encourages the Authors of submitted papers who are not English native speakers, to use a language service checking the language correctness not only with respect to grammar, but also in the way of presentation of research results accepted by renowned publishers, e.g. presented on the website of the European Association of Science Editors. The Editor encourages the Authors of submitted papers who are not English native speakers, to use a language service checking the language correctness not only with respect to grammar, but also in the way of presentation of research results accepted by renowned publishers, e.g. presented on the website of the European Association of Science Editors.


Figures
Figures (illustrations, photographs) and tables, provided in the camera-ready form suitable for reproduction (which may include reduction), should be additionally submitted (one per page), larger than the final size. While preparing figures we encourage to start with defining expected size and minimum font size that fit to all graphics in the manuscript – using the same style in all of your graphics visually improves the article. Final figure formats must be in one of the following: (vectors) .eps, .pdf, .ai or .cdr, and (bitmaps) .bmp, .gif, .tif or .jpg.
As far as plots, block diagrams, schematics etc. are concerned, we suggest to use one of vector formats to improve quality and scalability. Figures in vector formats must be saved using RGB colours and with fully white background (0% K). Hidden layers are unacceptable. Minimum line thickness printed in a single colour is 0.25 pt (0.09 mm), and 1 pt (0.36 mm) when using more colours. Typically we suggest 0.2-0.5 mm but in particular cases the range 0.1–1.0 mm will be accepted. Lines in plots should be distinguished not only by using different colours but also using different line types and markers, if needed.


Equation
All equations must be numbered consecutively throughout the text. Each equation should be preceded and followed by a 6-point spacing. Punctuate equations when they are part of a sentence. Equation numbers should be enclosed in parentheses. Equations should be prepared with the use of MathType or Microsoft Equation editors. The type size in the equation is the same as for the text. To make your equations more compact, you may use the appropriate mathematical symbols or expressions. The symbols used in an equation have to be defined before that equation or immediately after it. Use italics for variables (e.g. i, x, n), physical quantity symbol (e.g. voltage U, temperature T), letter pointers and general function symbols. Do not use italics for constants, indexes, minimum, maximum and trigonometric functions, mathematical operators, differentials, etc. To refer to the equation use “(1)”, not “Eq. (1)” or “equation (1)”, except at the beginning of a sentence where “Equation (1)” should be used. We recommend to use International System of Units SI i.e. metre-kilogram-second system of units. As a decimal separator dot should be used in the entire manuscript (text, figures, tables).


References
The paper has to be clearly positioned in the context of relevant literature in the field of measurements and instrumentation. Note that lack of references from the main field of Metrology and Measurement Systems interest may suggest that the content of manuscript does not exactly correspond to the scope of metrological journals. It may reduce possibility that a proposed paper will be read by audience society. In such a case our Editorial Board may suggest to send the manuscript to a more appropriate journal. Also note that the use of possibly up-to-date references may indicate importance of your work. Table below gives examples of some relevant and renewable journals related to widely understood metrology.


Journal

Publisher

ISSN

Metrologia

IOP Publishing

0026-1394

IEEE Transactions on Instrumentation and Measurement

IEEE

0018-9456

Measurement

Elsevier

0263-2241

Measurement Science and Technology

IOP Publishing

0957-0233

Metrology and Measurement Systems

PAS

0860-8229

Review of Scientific Instruments

IOP Publishing

0034-6748

IEEE Transactions on Industrial Electronics

IEEE

1557-9948

IET Science, Measurement & Technology

IET

1751-8822

Journal of Instrumentation

SISSA, IOP Publishing

1748-0221

Measurement Science Review

Walter de Gruyter

1335-8871

IEEE Instrumentation and Measurement Magazine

IEEE

1094-6969

Bulletin of the Polish Academy of Sciences: Technical Sciences

PAS

2300-1917

Opto-Electronics Review

PAS

1896-3757

IEEE Sensors Journal

IEEE

1558-1748

Sensors

MDPI

1424-8220




References should be inserted in the text in square brackets, i.e. [1]; their list, numbered in citation order, should appear at the end of the manuscript. The format of the references should follow the APA 7th edition formatting style, i.e.: for an journal paper – surname(s) and initial(s) of author(s), year in brackets, title of the paper, full journal name, volume, issue (in brackets) and page numbers. Put all author names unless there are more than 20. Otherwise, after the first 19 authors’ names, use an ellipsis in place of the remaining author names. Then, end with the final author’s name (do not place an ampersand before it).


Submission process
Manuscript should be submitted via the Internet Editorial System (IES) – an online submission and peer review system. In order to submit the manuscript via the IES, the authors (first-time users) must create an author account to obtain a user ID and password required to enter the system. The submission of the manuscript in a single file, i.e. “Article File” containing the complete manuscript (with all figures of high quality and tables embedded in the text), is preferred. All figures have to be uploaded in separate files. The generated PDF file has to be approved. The PDF file has lower quality of the embedded figures to limit its size only.
The submission of a manuscript means that its content has not been published previously, it is not under consideration for publication elsewhere, and that – if accepted – it will not be published elsewhere. The Author hereby grants the Polish Academy of Sciences (the Journal Owner) the license for commercial use of the article according to the Open Access License ( CC BY-NC-ND 4.0), which has to be signed before publication. The copyright form is available in the IES.
The Authors are urged to suggest 4 to 5 reviewers in their application (with names, affiliations and addresses) with whom the Editorial Board could co-operate while processing the paper. Proposed reviewers should be experts deeply involved in issues related to the subject matter of the paper and they are intended to come from different universities or research centres.
Each submitted manuscript is subject to a single-blind peer-review procedure, and the publication decision is based on the reviewers’ comments. If necessary, the authors may be invited to revise their manuscripts. On acceptance, manuscripts are subject to editorial amendment to exactly fit the journal style.
An essential criterion for the evaluation of submitted manuscripts is their potential impact on the research field, measured by the number of repeated quotations. Such papers are preferred at the evaluation and publication stages.
Proofs will be sent to the corresponding author by e-mail and should be returned within 48 hours from receipt. The publication in the journal is free of charge. A sample copy of the journal will be sent to the corresponding author free of charge. For colour pages the authors will be charged at the rate of 160 PLN or 80 EUR per page. The payment to the bank account of the main distributor (given in “Subscription Information”) must be completed before the date indicated by the Editorial Office.


Other information
It is possible to include supplementary files related to the article content, such as e.g. developed databases. These files can be then used by other researchers to compare their algorithms using the same input data. For more details about supplementary files please contact the Editorial Board: metrology@wat.edu.pl. The biographical statements, at the very end of the article, are not obligatory, however, they are kindly recommended. Each statement should include the author’s full name and brief personal history focused on areas of research and scientific achievements. The biographical statement may not exceed 100 words and should be written using Times New Roman style with a font size of 8 pt.
The publication of your article is a great achievement but then it needs to be further promoted to make it more visible to the research community. Responsibility for this task lies with the Authors and our Editorial Board. We guarantee free access to the article in the Journals PAN of the Polish Academy of Science, including articles in Early Access form (published just after acceptance decision), indexing in popular and renewable databases (e.g. Thomson Scientific Master Journal List, Elsevier’s Scopus, Google Scholar). Furthermore, selected articles are highlighted on the journal website and are reprinted for promotion at conferences and other events. The Authors can share the final form of the article on various social networks and research-sharing platforms, such as Twitter, Facebook, Linkedin, ResearchGate, Academia.edu, SciProfiles. They are also encouraged to update personal and institutional webpages by adding the title and a link of the article. Feel free also to share your work with your colleagues using any other methods that do not conflict with the CC BY-NC-ND 4.0 license.
For more detailed description about how to write a paper for the Metrology and Measurement Systems journal please look at the Author guidelines for manuscript preparation. We strongly recommend using this file as a template for manuscript preparation.


This page uses 'cookies'. Learn more